Fully Automatic Chemistry Analyzer
Technical specificationsGeneral informationMachine type Random access, open reagent & locked reagent (optional)Test speed Constant 120 tests/hour (Mono reagent or double reagent)Test principle Colorimetric method, turbidimetryTest method 1 point end, 2 point end, Fixed-Time, KineticsCalibrationtype Linear & nonlinear Reagent unitReagenttray 80 reagent positions (40 for R1 & 40 for R2), Non-stop cooling system with pelter pad inside, 24 hours 2C-14T, open system, could accept any reagent.Reagent volume 20-300y, step by 1lReagentneedle Liquid surface detection, timely tracking for liquid volume.Collision avoidance.Regent barcode reader (optional) Reaction unitReactiontray SOreaction cuvettes, new amorphous optical plastic reaction cups, perfect transmittance.Reaction volume 150,4-3301Reaction time 10 minutes Temperature Peltier pad incubation system, reaction temperature(37T £01T) Mixing systemMixer ‘pacdle-type mixing, effectively reducing o